10 Types of Records ISO 17025 Requires Your Lab to Keep
Why Records Are Central to ISO 17025
ISO/IEC 17025:2017 §7.11 requires laboratories to keep sufficient records to demonstrate that activities are carried out in accordance with the QMS and that results are traceable. These records must be demonstrably free from retrospective alteration — the main reason Excel without an audit trail cannot fully satisfy this requirement.
1. Raw Measurement Data
Original readings before any calculation or correction. §7.10 requires raw data to be stored separately from the final result, in a form that demonstrates it hasn't been modified.
2. Calibration Certificates
Certificates issued to customers must include all fields specified in §7.8.4: lab name/address, issue date, item description, method, results, and uncertainty. Superseded certificate versions must be traceable.
3. Equipment Records (§6.4.13)
For each instrument:
- Identity, make/model, manufacturer
- Date received and placed in service
- Full calibration history, with all certificates
- Maintenance and repair history
- Damage, malfunction, modification, or withdrawal from service
4. Reference Standard Calibration Records
Calibration certificates for all reference standards, with an unbroken traceability chain to NIMT (or another NMI), accessible at all times.
5. Uncertainty Budget Records
Uncertainty budget documentation for each method and calibration point: input data, assumed distributions, sensitivity coefficients, and results. Changes to the budget require a revision trail.
6. Internal Audit Records (§8.8)
Audit plan, checklist, nonconformities found, corrective actions taken, and objective evidence that each action was effective.
7. Management Review Records (§8.9)
Minutes of each management review: attendees, agenda items, QMS performance indicators, decisions, and action items.
8. Proficiency Testing Records (§7.7)
PT and ILC results, including z-scores or En numbers. Any failed PT must have a documented root cause analysis and corrective action.
9. Nonconformity and CAPA Records (§8.7)
Every nonconformity — from customer complaints, internal audit, PT, or self-detection — must be recorded with date, description, root cause, corrective action, due date, and effectiveness verification.
10. Personnel Training and Competence Records (§6.2)
Qualifications, training records, competency assessments, and authorisation records for each staff member whose work affects result validity.
Minimum Retention Periods
- Calibration certificates and raw data: 7–10 years
- Equipment records: full service life + 5 years after disposal
- QMS records (audit, management review, CAPA): 5 years
- Training records: employment duration + 5 years
What a Real Audit Trail Looks Like
In the context of ISO 17025 and 21 CFR Part 11, an audit trail records who did what, when, and what data changed — and cannot be deleted or retroactively modified.
LabSync uses a hash-chained, append-only audit log covering every action in the system — calibration result changes, job status transitions, certificate issuance — meeting ISO 17025 §7.11 and 21 CFR Part 11. Try LabSync free for 14 days and see your lab's record status on the compliance dashboard.